The members of the Solder Paste Working Group led by Dr Wallace Rubin, did
the testing that provided the needed R & r input data on test methods to the ISO
TC44/SC12 "Soldering Materials." In particular, we were evaluating the
Brookfield Viscometer and the Malcom Spiral Pump viscosity tester. In this case,
one of each of the instruments was circulated in turn to many laboratories. A
large sample of high viscosity paste was prepared by one company, a large
sample of low viscosity paste was prepared by a second company. These large samples
were sub-divided into small samples; then one each sample of high and low
viscosity pastes dispatched to each test laboratory. (As I recall, the two makers
were Alpha and Kester).
When all the data were available, they were evaluated by a Ministry of
Defence computer program to yield the needed 'R and r' results.
You may wish to contact Dr Rubin at [log in to unmask] or Malcom's
Tom Fujikawa at [log in to unmask] for further information.
Bill Kenyon
Global Centre Consulting
3336 Birmingham Drive
Fort Collins, CO 80526-2336
Tel 970.207.9586 Cell 970.980.6373
---------------------------------------------------
Technet Mail List provided as a service by IPC using LISTSERV 1.8e
To unsubscribe, send a message to [log in to unmask] with following text in
the BODY (NOT the subject field): SIGNOFF Technet
To temporarily halt or (re-start) delivery of Technet send e-mail to [log in to unmask]: SET Technet NOMAIL or (MAIL)
To receive ONE mailing per day of all the posts: send e-mail to [log in to unmask]: SET Technet Digest
Search the archives of previous posts at: http://listserv.ipc.org/archives
Please visit IPC web site http://www.ipc.org/contentpage.asp?Pageid=4.3.16 for additional information, or contact Keach Sasamori at [log in to unmask] or 847-615-7100 ext.2815
-----------------------------------------------------
|
|
|